Section: V1M2, Section 5.5.13.1.e.ii


REQUEST:

This section requires that disposable or single-use volumetric equipment shall be verified once per lot, prior to or in conjunction with first use.

Is it acceptable to perform this verification at one location of a network laboratory, and share that verification documentation among its laboratories, providing the required documentation of verification to any network laboratory that uses the lot in question?

TNI FINAL RESPONSE:

No, each accredited laboratory shall verify disposable or single-use volumetric devices once per lot, prior to or in conjunction with its first use.