Standard Interpretation
Standard: 2016
Section: V1M2, Section 5.5.13.1
REQUEST:
d) Temperature measuring devices shall be calibrated or verified at least annually. Calibration or verification shall be performed using a recognized National Metrology Institute traceable reference, such as NIST, when available.
Question: do reference thermometers need to be calibrated annually? These are traceable to NIST.
TNI FINAL RESPONSE:
As long as the reference thermometer is not being used as a piece of support equipment such as described in 5.5.13.1 c, a reference thermometer is a reference standard as described in 5.6.3.1 and would need to be calibrated before and after any adjustment, and in accordance with the lab’s documented procedure.

develop consensus standards for use by TNI's programs. This group has a support role in assisting other programs with activities such as guidance and standards interpretation.
fosters the generation of data of known and documented quality is through the National Environmental Laboratory Accreditation Program, or NELAP. The purpose of this program is to establish and implement a program for the accreditation of environmental laboratories.
is to establish and maintain a national PT program to support TNI’s National Environmental Laboratory Accreditation Program, and other activities.
Field Activities Program (NEFAP) Executive Committee is to establish and implement an accreditation program for field sampling and measurement organizations (FSMOs).
on a variety of topics of interest to the environmental measurement community. TNI also hosts two regularly scheduled meetings per year, typically in January and August. Both are a week long and involve open meetings of TNI committees as well as training courses and other special sessions.