Combined Interpretations of the 2003, 2009, and 2016 Standards that apply to Volume 1 of the 2016 TNI Standard
Question: I am having difficult interpreting the requirements outlined in 4.0. The main concern is with our metals department where we run methods 200.7, 6010B, 200.8, 6020. If we are analyzing a PT by all four methods and reporting all methods individually, are 200.7/6010B and 200.8/6020 being treated the same? For example, is a failure for Cobalt by 200.8 equivalent to a failure for Cobalt by 6020, even if our PT demonstrates that we passed Co by 6020? These methods have different digestions and different method requirement at the instrument level. For the 200 series we utilize a hot block digestion and the 6000 series utilizes a microwave digestion. At the instrument level, the control limits for MS/MSDs and blank spikes are different. The requirements for same-source and second-source checks are different. These are different methods. Is each metals failure for ICP a failure for all ICP methods and each ICP-MS failure a failure for all ICP-MS methods? If this is the case, are we able to only run by one method and hold the accreditation for both. The standard references FoPT, with is defined by matrix, technology/METHOD, analyte. Not just based on matrix, technology, analyte. TNI Response: The use of the term “method” within the definition of Field of Proficiency Testing (FoPT) (2009 V1M1, 3.6) is only included to accommodate EPA’s drinking water program where PTs are required per method for the drinking water analytes referenced in the Code of Federal Regulations (CFR), specifically 40 CFR 141. The use of the term “technology” within the definition of FoPT (2009 V1M1, 3.6) only refers to the determinative analytical technology; preparative techniques/methods are not part of this definition. In addition, the Note in Section 5.1.1 of V1M1, states the following: “…If the laboratory is accredited for multiple test methods that use the same technology within a field of accreditation, the laboratory is not required to analyze a PT sample for each test method, except for fields of accreditation for the drinking water accreditation matrix for which a PT sample per test method is required…” Therefore, using the example provided, for each analyte in the same matrix, the TNI standard only requires PTs for one ICP method (200.7 or 6010B) to maintain accreditation for both ICP methods and one ICP-MS method (200.8 or 6020) to maintain accreditation for both ICP-MS methods. V1M1 of the 2016 standard was revised to include this statement "An unacceptable score for the reported test method will result in an unacceptable score for all test methods for that accreditation FoPT. Question: A laboratory in our program has requested accreditation to measure analytes in biological tissue. The question is "If biological tissues are not listed as a matrix for the current NELAC Fields of Proficiency Testing, are proficiency tests of solid and chemical materials acceptable to demonstrate proficiency for testing biological testing?" TNI Response: Biological tissues are not a matrix in the TNI FoPT tables, as such there would be no proficiency testing requirements for this matrix. The 2016 standard clearly indicates only analytes in FoPT tables are required.
MODULE 1: PROFICIENCY TESTING REQUIREMENTS
Section: 4.1